User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2554 VERIFICATION PROCEDURE (ver1.0) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 45 NUMBER OF LINES: 87 CONFIGURATION: Fluke 5700A (WB) ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Auto Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P A W 1.002 HEAD {VOLTAGE PERFORMANCE VERIFICATION VIA WIDE BAND} 1.003 DISP CONNECT THE PM 2554 TO THE 5700 BY MEANS OF A BNC CABLE, 1.003 DISP AND A 50 OHM TERMINATOR AS CLOSE AS POSSIBLE TO THE UUT. 1.003 DISP VERIFY THAT THE UUT HAS BEEN WARMED UP FOR HALF AN HOUR. 1.003 DISP CHECK THE RESPONSE SWITCH BEFORE VERIFYING THE UUT 1.004 5700 1mV 2% 10H W 2W 2.001 5700 1mV 2% 100kH W 2W 3.001 5700 1mV 3% 1MH W 2W 4.001 5700 1mV 5% 5MH W 2W 5.001 5700 1mV 8% 12MH W 2W 6.001 5700 3mV 2% 10H W 2W 7.001 5700 3mV 2% 100kH W 2W 8.001 5700 3mV 3% 1MH W 2W 9.001 5700 3mV 5% 5MH W 2W 10.001 5700 3mV 8% 12MH W 2W 11.001 5700 10mV 2% 10H W 2W 12.001 5700 10mV 2% 100kH W 2W 13.001 5700 10mV 3% 1MH W 2W 14.001 5700 10mV 5% 5MH W 2W 15.001 5700 10mV 8% 12MH W 2W 16.001 5700 30mV 2% 10H W 2W 17.001 5700 30mV 2% 100kH W 2W 18.001 5700 30mV 3% 1MH W 2W 19.001 5700 30mV 5% 5MH W 2W 20.001 5700 30mV 8% 12MH W 2W 21.001 5700 100mV 2% 10H W 2W 22.001 5700 100mV 2% 100kH W 2W 23.001 5700 100mV 3% 1MH W 2W 24.001 5700 100mV 5% 5MH W 2W 25.001 5700 100mV 8% 12MH W 2W 26.001 5700 300mV 2% 10H W 2W 27.001 5700 300mV 2% 100kH W 2W 28.001 5700 300mV 3% 1MH W 2W 29.001 5700 300mV 5% 5MH W 2W 30.001 5700 300mV 8% 12MH W 2W 31.001 5700 1V 2% 10H W 2W 32.001 5700 1V 2% 100kH W 2W 33.001 5700 1V 3% 1MH W 2W 34.001 5700 1V 5% 5MH W 2W 35.001 5700 1V 8% 12MH W 2W 36.001 5700 3V 2% 10H W 2W 37.001 5700 3V 2% 100kH W 2W 38.001 5700 3V 3% 1MH W 2W 39.001 5700 3V 5% 5MH W 2W 40.001 5700 3V 8% 12MH W 2W 41.001 HEAD {ALTERNATING HIGH VOLTAGE PERFORMANCE VERIFICATION} 41.002 DISP [27][91]5m NEW CONNECTION INSTRUCTION FOLLOWS 41.003 DISP [D500][007][D500][007][D500][007][D500][007][D500][007] 41.003 DISP REMOVE THE BNC CABLE AND TERMINATOR 41.003 DISP CONNECT THE UUT TO THE OUTPUT TERMINALS OF THE 5700 41.004 5700 10V 2% 1000H 2W 42.001 5700 30V 2% 1000H 2W 43.001 5700 100V 2% 1000H 2W 44.001 5700 300V 2% 1000H 2W 45.001 END